<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\NVSStyleSheet.xsl"?>
<Parameter>
<ParameterFamily>al</ParameterFamily>
<LongName>
external_alarm_mode
</LongName>
<Name>
xamode
</Name>
<AppliesTo>
<UnnumberedList>
<ListItem><TextApplies>ThermoFlex</TextApplies></ListItem>
</UnnumberedList>
</AppliesTo>
<Description>
<DescriptionText><text>
The machines with this parameter support an external alarm of the customers.
This parameter sets when the alarm should be activated.  The following table
describes which setting does what:
</text></DescriptionText>
<DescriptionText><text>
xamode  |L |U |E |
</text></DescriptionText>
<DescriptionText><UnnumberedList>
<ListItem><Text>   0  | - | - | - |</Text></ListItem>
<ListItem><Text>   1  | x | - | - |</Text></ListItem>
<ListItem><Text>   2  | - | x | - |</Text></ListItem>
<ListItem><Text>   3  | x | x | - |</Text></ListItem>
<ListItem><Text>   4  | - | - | x |</Text></ListItem>
<ListItem><Text>   5  | x | - | x |</Text></ListItem>
<ListItem><Text>   6  | - | x | x |</Text></ListItem>
<ListItem><Text>   7  | x | x | x |</Text></ListItem>
</UnnumberedList></DescriptionText>
<DescriptionText><text>
The columns L, U and E correspond to Load, Unload and Error respectively and indicate during
which states the alarm will be operated.  Note that combinations are possible.
Where an x is present the alarm will be active, a - means no alarm.
</text></DescriptionText>
<DescriptionText><text>
In each case, the alarm will remain on until either the error is cleared or a sequence is
started.
</text></DescriptionText>
</Description>
<Type>
uint16
</Type>
<Default>
<UnnumberedList>
<li>0</li>
</UnnumberedList>
</Default>
<Range>
<UnnumberedList>
<li>0 to 255 (no units)</li>
</UnnumberedList>
</Range>
<ArraySize>
</ArraySize>
<Notes>
<DescriptionText><text>
</text></DescriptionText>
</Notes>
<Partition>
cal
</Partition>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<li><Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link></li>
<li><Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link></li>
<li><Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link></li>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:35:44 2008 PST
</Updated>
<Links>
</Links>
</Parameter>